Hello uh this video it's uh for uh Nanos 3 Talos f200x temm training my name is gfu I am the process engineer at Lando stre facility ussd um so first one it's about our uh TM thof fure Talos f200 XS TM uh for this TM we have 200 KV fakeer can do high resolution TM Imaging um also equipped with scanning transmission electron microscope STM with bright field dark field and the high angle angular dark field detectors equipped with super X4 sdd energy dispersive xray spectroscope edx system and the electron energy lost spectral scope used by
using Gens um spectrometer with utra fast du use Spectrum du use means can simultaneously acquire L loss and the core lost uh Spectrum for this training we will cover following uh contents the first one it's mounting sample to single teed holder and double teed holder the second it's step the second training it's procedure to insert the single or double TI holder is the TM then uh we will do the TM Alignment and the Imaging after that it's STM alignment and the Imaging um then we will cover the STM energy dispersive X3 spec scopy edx analysis
um then it's a temm defraction alignment and image this is the first one mounting sample to single TI holder and the Double T holder we will uh step by step show users how to mount sample to single To and a Double T holder after sample mounted to single or double til holder we we will insert the holder in maxos scope I will demonstrate the procedures to insert the single TI holder or double TI holder in Max scope our TM have a touch screen near the computer near the TM so I will demonstrate how to use
touchcreen to finish the insert Holder process in the same time some users want to use the interface of main software I will also demonstrate the process to insert the holder by using the main software user interface for TM alignment and imaging I will using the cross grading TM sample to demonstrate how we open the beam find the sample doing the sample entric alignment and the alignment the beam Alignment by using director alignment I will show the steps by step to finish the TM alignment and the get TM Imaging for STM I will demonstrate this function
by using a larer sample uh this is semiconductor TM sample um we will uh showing how to get uh by using different detector bright field dark field high angle angular dark field detector to get the features different um features by same location For edx application we will using the um Super X edx system to looking for the same semiconductor sample and find out one interesting area for example in this picture we can go to the transor area and doing edx mapping so to find out different materials there Like Oxygen silicon nitr and the co Cobalt
for the single Crystal samples for example cicon I also will demonstrate how to use temm to doing Deflection uh to align the sample on the room axis so the beam can pan penetrate through the room access and also to do the Imaging for the single Crystal samples yeah this is the uh contents for this training yeah let's start for the temm they have two type of the sample one it's 3 mm this kind of sample uh like the round sample and uh uh this sample have two side one side it's a little Darker uh that
one US is carbon fume so uh if the sample is the Nano particles Nano wire you want to put your sample on the darker side which is on the carbon fume okay uh and we use this sample as as a example and uh for using the single T holder okay so when get the holder the holder we put in this location it's safe and when we need to mount the sample we move the holder to here okay so this is the single til holder and uh uh the holder looks like That so we have I
have glove and I can touch this side don't touch this side even have glove so this is the O-ring to uh for the vacuum and here have a pin there and the sample mounting on the top okay and I will be put it in doing Mount the sample I open the top part and they have a small sharp leadle this leadle can be put on this side uh but if you use left hand right hand Some users put the needle on other side okay so just depends on what is last user they handled the the
um used the holder okay so right now I move the holder to this way this is Lally I mount the sample okay so the needle they have a small hole here and I just inserted here and uh my right hand to hold this on because I don't want to move okay hold this side and I lift this Lever okay gently to all the way to Vertical like this okay then they open the space I can mount the sample on there so just like I mentioned the sample have two side one side is like the copper
color other side is a little darker which have the carbon F when we mounted we want to the carbon side to be facing down okay then in the microscope since this Holder will turn 180° then in the microscope the coer carbon side will be facing up but right Now we mounted facing down for the carbon side I opened the box okay so uh when handling the sample um I prefer use vacuum traser instead of the normal traser because Normal traser sometimes still easy to damage the sample so when touch the sample I'm using the vacuum
traser this is the vacuum traser just the one button open close okay when it's open the they start to have the vacuum so they have a hole here when you use the finger to Block that hole then I can lift the sample out something like that okay and when finish I just release the finger then the sample will be dropped okay I do one more time close the finger get the sample up and open the finger to releas the sample okay so right now I will Mount the sample on the single TI holder okay so
right now the sample is it's There then want to verify the location make sure it's on the right position right now it's pretty good okay then I use the needle again right hand hold it make sure the holder not move and the left hand I get this needle get this hole and slowly get the lever done okay and make sure it's all the way done so this is how we Mount the sample on the single T holder okay then I tiled this one and give a little tab make sure the sample is safely mounted on
the holder Okay to get the sample out when you finish the experiment you do the same thing okay so use this one insert this small hole and the right hand to to hold this one and lift the lever all the way up do not leave halfway okay because halfway they may come back by the spring that is sometimes it's dangerous so just all the way up then I can use vacuum tweezer to get the sample out of the holder and back to the Box Then hold this side and slowly makes the lever down and sample
save back to the box and close this one so this is the process mounting sample and the D mounting sample for the single TI holder okay so uh let's go to the double to holder so we put this one back to here to be more safe then I get the Double T holder uh um a aot a little thing to explain so what means single TI what it Means Double T single TI means when the holder inserted in the TM column the holder itself cannot move itself they will the the til with a stage so
the holder is inserted the T with a stage so this means single til Double T this one the difference with single Tu with double Tu they have a cable okay this this cable is connected to a connector on the TM we will show you later and the holder itself also can go with the stage to tilt this is Alpha tilt okay also Like a sing Single til holder only can do alpha TI but for the double til holder they have a cable this cable have the power to drive a lot a small function on the
top so on the top they can do not only Alpha they also can do beta so means two directions til Alpha and beta so this is why they call the double til holder okay so for this holder looks like other part looks like seminar so uh wear glove will touch this kind of holder and this is O-ring do not attach anything above the O-ring to the top and also they have a pin there this pin will be used during the sample insertion okay so let's do with the sample mounting process okay so um this needle
is only for single til holder so I put it back for double til holder we need a special small needle okay I get the little Out so this needle it's a it's sharper and a very small little so this one will be using for Mount the uh sample on the double tued holder okay uh for double toed holder we will use different sample not this one we use different sample okay uh for the single tud holder you know that sample is around 3 mm for this one we will do the different sample I open It
so can take a look okay so for this one instead of the circle round sample this one is more like a half moon or it's also 3 mm but it's like a half moon sample so this is a different TM sample this is uh this sample is uh made by using the focus sign Bean uh system U do bean Focus sign be system um before Mount the sample let's look at the how we hold the sample there Okay I move this one a little away so can see here so right now from here you can
see they have a coer clap or something like that when need to mount the sample we turn the holder to the back okay then from there we can see a very small hole on the top this is I marked here so so I need to use this needle into that small hole and push okay so right now you'll say two things stop one is the Spacer why is the spacer one is a clamp okay so this two will be used to mount the sample right now I turn this back and the you can see original
the clamp is on this position and right now from the back I push by using needle the clamp and the spacer job okay and right now I can mount the sample on the Double T holder okay and here just like I mentioned we have two samples there you can see this uh like a half moon and also you can see one It's corer color another one is like the server color I just want to show uh we also have different materials for the sample base like this is copper this is m o why we need
to have different one because you know the tm1 the function is e DX by using elect be to figure out what is the materials so if your materials um have the coer then you don't want to the base use copper you want to base use other materials otherwise when the edx show The Copper Signal you don't know it's a signal from the sample or from the base from the grid okay uh the same thing for for another one okay so right now I will Mount one of them on it I also use vacuum traser same
process I get the sample I put the sample on the holder and I make sure the sample it's on the the right position okay sometimes I just use a little tab to let them go To the position okay then after that I can put the spacer on it okay when get the spacer put on it um this I can use the physic Tracer and do not grab spacer this way this way may the spacer fly away so grab like a horizontal like this way and I put my spacer on it okay then this is a
clamp first I need to make sure you can see the clamp there they have the SC right now is Facing up I need to make sure the clamp facing down and uh for this clamp you can see see this two is closer compared to this one so this two will be on here and this two will be on this side so I mark it the right direction then I put this one on okay then I use small twers to adjust this two to make sure this two on the right position and also this two right
now is clamped down so this two already on the right position and clamp down but For the top right now it's still up so I will use the tweezers to push the top to finish the clamp process and when use the tweezers make sure don't do this way this way will be damaged the sample use this flat side okay I will do the demo right now okay I heard some click sound there then it's the clamp is on right position then I turn this one 180° a little tab make sure the Sample it's not Dro
or make sure the Moun is good okay so this is how we Mount samples on the double til holder okay to get the sample out when you finish the TM turn back use this small little find the hole there and pushing okay okay so right now you can see three s stop clamp spacer and the sample right So use vacuum tweezer for the sample okay sample is safe and uh I put the spacer on it right now low sample but you need to put the everything back okay then the next step is for the clamp
find out the clamp the direction then put this one on top okay same thing you this one to push this side the clamp down to finish to Put the clamp on the right position okay then this is get the everything back so when finish double TI holder both hand to move it back make sure use both hand because this one it's May slides out and jop okay make sure carefully put it to here this needle put it into the box and save here tracers okay so this is how we put in The sample on single
T holder and double til holder so um for the single til holder um before insert this is the touch screen so we need to focus on two things the first one is this valve this one this one is close the second one is the stage is all zero if the stage number is not zero so reset XYZ to make it zero okay so right now insert the single TI holder so uh usually this door is closed like this so uh we open the door and and uh we get the holder so uh I have glove
holder here have the pin so we insert the holder okay just like that you see you hear the sound that one is the reading so right now they start to pump so and also you see the red LED it's on so that means right now they start to pump down we need to wait this is a 4 minute 4 minutes process wait the pump down process to finish when they pump down pre pump the finish this light will be turned off when the uh single T Holder inserted to the uh loading area they will jump
up a window ask you to select what kind of holder it is so right now they should Double T holder because previous user you use double since right now we use single holder so we need to from this menu to select the single TI holder okay after we select the right holder then we click select okay um this one it's uh for the for the pre Pump we sit like 4 minutes so right now they have still like 20 three 3 minutes uh 10 seconds for the pump down after the pump down the process we
can fully insert so right now we wait this one to be fully finished during the pump down process uh on the sample the light also it's on LED light so we wait uh another 2 3 minutes okay so finished the uh pre pumping and this is the lmal okay so this is the caution this is okay if they have some Vacuum leak or something a a a window will jump out and say tuber pump error or something right now this is good so I can finish the inserting process so right now I will counter cross
wise turn around 120° this big pin will into this hole the vacuum will suck this holder in I just hold it let the smoothly come in okay I use this one gently click this one to release something like the oing the tension or something so make the Holder to be in the microscope more stable not drift okay so this is insert process okay so right now I finish the inserting process you can see the colum vacuum before they jump to like 16 but right now they jump to uh they go down like uh 13 12
11 and 10 so this a good sign so if this colum vacum increase and keep like the high level like 30 35 then we need to continue waiting till this vacuum going down okay uh so this is inserting process when we want to get The sample holder out uh for the touch screen or we also need to check two things the first one make sure this valve is closed the second one is reset the stage X and Y XYZ Alpha and beta so click this one they will reset the stage to zero then um I'm
good to uh retract the holder we get the holder out we do three steps first one it's just pull the holder then we will fill the stop then we will do the clockwise Turn then we fill the Stop and the pull again so three steps okay so first one turn second one third one so this is for the single til holder right now I will uh demonstrate for the double TI holder the process so uh right now before inserting double check make sure the valve is closed and also reset the stage okay so it's ready
for insert this is a double te holder so Compared to the single TI holder uh they just have one cable and this cable need to be go to this connector so um the process we're doing the process uh stage it's on the right position and the vacuum is closed so this is a pin and we insert from here okay so right now this night is on so they start to pump down the first step is the same like a single T holder we Insert the double te holder in the um uh pre pump area uh
they also have a window jump out ask you to select the holder type since last time we use single to holder So Right Now the default is single to holder from here we can go in open the touch screen and see the menu right now we are doing the double Ted holder so we choose double Ted holder then we click select Okay so for the double til holder we also need to connect the holder cable before the cable connect you cannot confirm this one so right now I can connect the cable so the cable is
for this connector um the top part is a little wider so this part is a little wider so it's not like a square yeah so this one I open it and insert I already connect the cable uh From the holder to the stage so right now this one become to available and I can can confirm right now the cable is good connected we continue to wait another 2 minutes 30 seconds to finish the pre pump process then we can fully insert the holder okay so night is off and this is a normal if you have
any Sy wrong our window will jump out say some Tu pump issue or the Vacuum issue right now this is good so with um counterclockwise uh turn 120° to finish the insert process counterclockwise uh 120° then the vacuum will suck the holder in release some tension okay so this is ready for analysis okay um before analysis make sure close the door okay so this is good You can see the vacuum previous day 21 then 15 11 10 they continue going down so this a good sign and Al right now it's nine so that means no
uh vacuum break and also the sample it's good it's don't outgasing to increase the vacuum okay so when we finished the um analysis by using double te holder before we retract the holder we need to double check make sure two two things the first one it's the valve we want to make sure the valve is closed The second one it's we also need to Res set XYZ Alpha Beta make sure uh the stage is on the loading position okay so right now we can start to retract the holder since this is it's double to holder
so they have a cable connect to the stage oh we to remove the cable first okay the first Ty thing um this one is after we verified the valve is closed and the stage is zero then we can remove the cable first okay then after the cable released This one become to light up that's that's fine because they just say okay they lost the connection here okay right now I removed the cable since they lost the connection so the software jump a window say oh what is the holder because they if it's double tiled they
cannot they they don't have any connectivity with the cable so they jump this one so this is okay then uh the rest of the step to remove the holder is the same like a single TI holder then for this One still three steps P we you feel the stop clockwise turn when you feel the Stop and pull again okay holder get out okay and close the door here okay so this is sample uh holder Double T Holder have been removed and then you can say the vacuum already back to one column vacuum that's good yeah
so that means everything good no vacuum leak and also from here you Can see XYZ and also you can see the liquid nitrogen level right now you see here it's 90% um so this is very almost the full of the liquid nitrogen if this one lower than 15% then um let the dinos stuff loow so we need to refill the liquid nitrogen okay uh I will uh start to loading sample uh before loading sample we want to make sure the column vacuum it's Small number like five or less than 10 KN and also we would
like to make sure the colum valve is closed the colum valve can be sa from the vacuum overview it's this w we want to be make sure it's closed when we insert the holder they have some small chance we break the vacuum or if this valve is closed then just affect the vacuum in some other camera area or the column and take like a half hour to recover but if this valve is open and break the Val Vum uh then the filament will be burned uh damage and take much longer to recover uh uh need
to replace that is two 3 weeks job okay so right now this valve is closed and uh we will also make sure XYZ Alpha Beta set to zero so I will go to the search reset holder okay to reset everything to zero also check objective and select error it's retracted okay so right now I will insert the Sample insert the holder I first I'm insert the single til holder from here you can see right now I inserted the holder to this uh uh uh loading knock or partial there and this start to pump down it
take like uh uh 4 minutes for pre pound then I fully insert the holder in the chamber so select the spacement holder this is the single Ted so I click single T and I confirm we also can see the air to air Lock the cycle time remaining it's around for three minutes while we are waiting for the um air log cycle time to finish we can go through some of the uh user interface so this part it's uh TM control they have different pages like a setup search alignment camera and a lot of sa there
okay and uh uh in each page for example in the first page they also have one two Three four four boxes for example in here set up we have the vacuum high tension F control and FG register okay on Lower uh Corner lower left corner this is um for hand panel you can see here the hand panel on the center part here this is multifunction X and Y L1 2 3 and R 1 2 3 what is the function for this lobs or buttons it's showing on here okay so for example currently multifunction X it's
beam shift X Multifunction y it's beam shift y this part it's the Noti ification area so if for anything like the sample uh insert the holder select the different sample that we showing on here this area it's tem right field uh magnication and mode right now this is uh previous the setting we will see this kind of change when the sample finish pump down okay 30 second we'll wait then we'll insert the fully insert the Sample for TM sample is fully inserted and uh then we can check the C vacuum right now it's 10 usually
we wait one or 2 minutes we want to make sure this L going down so for example right now this number become to9 that's good uh we do if this number going up we need to wait a little longer to make sure the lumber will be stable there or going down Okay if the L continue going up or they stay in some high L like 30 35 and don't go in down one of the reason is the due to the sample if the sample is not well prepared or not totally dry they may still out
gasing which make the chamber colum Vacuum cannot going down uh if that kind of case you need to remove the sample and the further uh dry the sample before put it back to temm okay so right now the lumber become to Eight it's continue going down so we can uh start to open the um open the valve okay so you can see the valve it's open okay after that we would like to load the file for the Fig register file we select the TM 200 KY spot five and click set and uh press this entric
Focus to reset some of the lens parameters okay right now you can see this Magnification it's pretty high so we want to lower magnication and expand the beam for TM bright field they have different mode to operate for example right now this is M mode if we continue going to low magnification we can say it into a mode okay so right now we can see a lot of mesches in this TM grid this black one is copper and this one it's carbon fume and this is the broken carbon fum Okay so I'm looking for maybe
this location uh I would like also like to use some of the vacuum area for alignment for the team most of the analysis most the users use analysis in mode with the magnification in the mode the smallest magnification it's 4,300 and the highest one it's 630,000 you can see that one I continue increase this is a high the sa magnication if you go to Le magnication It will be jump from sa to another MH mode uh since most of our manic uh alignment it's in sa mode so an sa mode cover pretty large range so
we would like to focus on this mode but if you need higher magnication you can go to the next level and just repeat the alignment okay okay I use joysticks to move to some areas where I mean features we can move to the center of the be frle here and the next step we need to do it's Doing some uh obler we use obler method to uh adjust the entric height uh here you can see uh right now the column continue going down that's good see I continue morning it monitor this number in the first
uh 10 minutes or something and another thing I also want to mention before we continue for the entric it's this lrien navel right now the lren nael is 29 9% uh usually uh this is for the liquid nitrogen the cooling down the cold Finger to keep the good vacuum of the of the column uh usually they consume the level is 1% 1 hour so right now 29% still pretty good for today uh next few hours no problem uh if you see this L below 15% please let the n three stuff low so we can refill
the liquid lri okay so for the entric we go to the search we do the alpha obl right now we can set the til degree from 0 to 15 usually uh start with 10 or8 uh it's Good number to start why click the opener you can see the screen they are moving the features they're moving means right now the screen height is not entric height so we will adjust Z to make it uh entric um minimize minimize this kind of move uh this Z it's a pressure sensitive uh button so you press and hold it
then it will be move and I can see the movement by here by the Z lber so now it's zero let's do some Changes okay so looks good right now you can see they are moving in a very small range the scale bar is one micron we would like it if moving within Micon or half Micon that is best so right now I think it's good it's already within one Micon so we can stop War a lot other method for doing the entric uh adjust entric we call it contrast method so right now I increase
the medication to a little high so we can see more clearly the contrast the Method uh it's I right now we want to the minimum contrast okay so right now I will adjust the Z so you can see the screen the change okay if I do this way you see oh the contrast increase you can see this features pretty clear if I do other side you see certainly the contrast it's pretty low minimum like here right now if I continue adjust Z you can see the contrast coming back so contrast good contrast High minimum contrast
contrast Higher again so we would like to the minimum contrast okay something like here so right now the sample is in your Centric height uh we finished the sample alignment then we need to align for the beam so usually uh we would like to go to the vacuum area for the Align for the beam so we decrease the medication we go to the vacuum area that means low sample if you cannot find the vacuum Area you also can do the beam alignment but the vacuum area will be a little easier so um okay this is
a vacuum area I use intensity to shrink the beam to a point right now you can see the P the beam Center beam it's not in the center of the foser screen so we go to here directly alignment and we use beam shift to move the beam from here to the center and before I click the beam shift you can see right now multifunction XY it's Beam shift but this is not alignment beam shift then when I come to here click beam shift you see the multif function X and Y they change it to alignment
beam shift X and Y so I can use multifunction l x and y we can move the beam to the center okay then we finish this one we click down so in this way we release the multif function from this alignment to their original settings okay then the next step is about the apure Uh for the aperture for example we can see from this side this aperture box our Max group have condensed aperture one condens aperture two and some other aperture and right now we just uh need to use condenser aperture to this one should
be always inserted the same with condenser uh aperture one for the objective aperture or selective aperture that is for some space specific application right now I don't need to use this two so this two right now it's Retracted okay so for the aperture we need to expand the beIN to the outside the red circle okay and right now you can see um right now this aperture it's a little to s clock or 8:00 Direction uh it's not fully in the center so we need to adjust the condenser to aperture right now uh after I clicked
the adjust the multifunction ex and Y we also showing adjust the condenser apure but right now I haven't adjusted it so it's still beam Shift X and Y so when I click it you can see multifunction X and Y Z they change it from original one to the apture X and apture Y so I can use these two to adjust make sure to adjust to Center this aperture okay something like that yeah uh just also want to let you low for this aperture right now I'm using 150 uh our system also have different size of
condenser tube aperture you can see from 150 100 70 and 20 the unit is Micron so for example I expand this being this is a 150 and when I change to 100 you'll see the some of this area is been blocked by the aperture only this size coming coming down and I can continue if we lead it to the smaller one like a 70 and uh 20 okay so each time when you change it you may also need to do a adjustment for the aperture so for example right now I go back to 150 I
still need to do a little bit adjust because right now this one is a little 1:00 2:00 Direction so I need to adjust a little bit for that okay looks good so when finished I click this one again so we can release the multif function X and Y from current one go back to their original setting beam shift X and Y we uh the next one is uh we want to Also make sure the beam it's round it's like a circle so we need to increase medication to a little high to verify this information right
now the medication is 7,000 so we adjust medication for example let's go to like a uh close to 100,000 okay something like that looks good if Le the step is 1 20 so yeah this one is close to 100,000 94,000 and you can see right now the Beam it's not in the center yeah and also you can see this one it's intensity it's very sensitive the reason is right now in this medication I want to like to find so I click here this light on so right now I can easily to adjust it I make
this one as a point you can see this point not in the center so I use beam shift to move the beam to the center Okay then I expand the beam okay looks pretty okay for me I adjust a little of the aperture do some fine adjustment okay and the beam it's pretty round but uh uh if it's not round we can always adjust to the Stig so come to here we see the stigma and we are using the condenser when I click this one you can see they have a different L there And this
number have a so this is a group one group two and a group three and group two you see they have a square there so that means this is the current setting before we make any changes and this is change steps before we make any changes we would like to keep the current l number in case we need to go back so how to keep this number right click and select so remember this is number one 2 three so current one right click and copy 2 to one so that means we Make a copy of
this number then we can make the changes for uh condenser because right now you say multifunction x y this condenser stigma X and Y so I can make some changes oh this is not good this is X uh this is x a l Direction and also you can change our number y yeah so make it R uh make it uh better so but if you if we messed up we say oh I think the original one is better since your original saved the data so so I can just click here and go Back original one
or sometimes you make something although looks like pretty good but you don't know oh which one is better so you can compare yeah to figure out which one is better when finish the stigma you can also click l to confirm this settings so right now I think right now this one it's good enough so I click this one to save the data the next alignment we would like to do uh we go back to the here directly Alignment it's the beam tiled PPX and beam tiled ppy when we want to do this we want to
uh mancation like over 120 sometimes 150 so we increase man ation and we shink the beam to a point okay and you know right now the beam is not in the center so I would like to use beam shift to move the beam to the center first in this magnication it's also pretty sensitive so I can use this two Find F cause to adjust this part the sensitivity okay so right now it's much easier to align it then finish the beam align uh beam shift and I click down then right now I use I am
working on the beam til PPX ppy why like the other functions when I click it the multif function X and y we're changing from current setting to beam til PPX and ppy okay so right now you can see two is Blinking we want to move the two together right now this two pretty close but sometimes it's a little far away so for example this is X and this is y okay sometimes you see the be looks like that so you need to adjust h and y to make these two Beam on the same place something
like that so this beam T PPX then do the beam T ppy the same example like a two beam sometimes like that then you need to adjust it to make Sure the two on the same location okay something like that then this is aligned PPX and PP y uh the beam a little not in the center so beam shift move the beam back to the center okay so the next alignment we should we will do is rotation Center for this one we need to find the sample right now you know we are moving to the
vacuum area so I will lower a little bit medication to find the sample and then move the sample to to here okay sample already in in the uh in the view so I increase medication for the rotation Center we would like to go to 120 or even higher medication okay so for example right now I go to like 150,000 meditation the screen it's so much Lowest you can see it like the signal lowest the ratio is not good when this kind of happen please use intensity to to shrink the beam so improve the signal yeah
for this magnication right now it's cost I adjust the to fine and also Beam not exactly in the center so I will move the beam shift to move the beam in the center okay expand the beam okay so we can see a bunch of the Features there that's good then when I click the rotation Center the Imaging is like room in room out basically the under Focus over Focus under Focus over Focus but uh right now they are doing like this but they also have a little X and Y movement I make it that one
a little more easy for your SE okay so right now for example you can see the IM image they're doing like this room in room out but they also do like this direction or sometimes you can see They also do y direction or some combine X and Y so we need to use multifunction X and Y to minimize this kind of move L than just focus on the focus or focus and focus and minimize this kind of XY movement here since I already select the rotation Center so you can see the multifunction X and Y
they already changed to rotation Center X and Y okay for okay so right now Focus under focus Photos Focus under Focus but the the the features not XY move that is what we want finish this one we click down to release m function X Y from rotation Center X Y to the original settings okay if we find out some features for example right now or some place you want to take some pictures we would like to use the camera so we go to the camera page and sometimes you'll say wow where is the camera looks
like the hardware not Connected okay so if you find out this issue double check is a camera software it's open or not the software for the camera they call it Tia T A so like this one so I need to open the software when the software is open you can see oh here the the the hardware uh seea camera is here and a bunch of a setting coming down so this is uh all the settings we may need to use to operate the Max uh camera but if you don't see this information just the Double
check it's a software open or not okay when we see the camera when you first use the camera you should see like this kind of thing if you see the insert is yellow that means the camera still inserted by previous the user and forgot to re uh retracted example if you see something like that that means the camera is inserted if you when you open your set like that so before we use the camera we want to set the parameter so if that kind of case I would uh suggest You remove the camera first check
the parameter make sure everything is good then reinsert the camera to check the parameter of the camera they have a small triangle here open this window and here this part it's search preview acquired we would like to check this parameter this param meter the main thing we are checking is the integration time because you know the electron being penetrated the sample and Coming down the camera just under it and the camera can only sustain certain level of electron touch their surface if too much electron coming down it may damage the camera so this is why
we want to control how much electron can reach the camera by setting the integration time so um let's take a look search integration time 0.128 second this is our preferred setting for the search and the preview The same number and the acquire we want to 1 second okay if you see some of the L it's not this number we would like to change to this number for example if you see something integration time is 4 second you can ler to 1 second or you can just use keyboard to type in 1 second but if you
change it by typing the L please click here to confirm this L otherwise the Lum not really input for example if I go back to here I didn't confirm for the acquire I go back you Say Ste 4 second you need to confirm the Lum then the Lum saved to the system okay in order to take pictures right now for example on the certain on this location you want to uh save the pictures so um first I set up the folder for the picture give me one second okay so um I click search and then
move okay when you click search if the camera is not inserted They will start to insert the camera then start to search but right now you see this pictures or this features not show on here the reason it's the FL camera FL uh forer screen is above the camera the real CCD camera and being coming down so I need to retract the forer screen or leave the screen so the uh Electron Beam can reach the camera so uh re uh retract the screen it's R1 you see here screen lift I click here then the features
go back to the Camera and right now I need to come to here I want to verify one important parameter it's you see my mouse it's here but I'm looking for this area when I Mouse move to the sample so you see the maximum number it's 18 1900 or something that is too high for the camera to sustain during the acquir this number should be lower than 1,000 counts so we go back to the fal first screen and you can see on the right they have a blue green bar the height means Current intensity right
now you say 1,800 1900 before we count it's on this range so we need to use intensity to lower this intensity the bar okay so we go okay this is wrong direction looks like go to higher okay we want to a little bit lower okay something like that then we open open here leave the that one and I want to recheck the counts okay so you see right now the maximum counts nowhere from 1800 1900 to 900 1,000 still still a little high so I want to lower a little more insert the screen this is
the height for that current intensity I want to lower a little more okay go back to the camera and uh looking for the maximum number Here okay so right now you see the max number it's 600 700 800 so they drop previously like 1,000 1800 so they drop to the label we want then we can start to take the picture by click acquire okay so this is the picture we just take uh to save this one we can file save as go to the folder go to computer user data uh select the folder you want
right now I Set have a training folder and this is the original file we would like to SA save okay so after the save you can see on here I saved one file but in fact it's one image but they saved two file this is the original file uh the original file it's specific file L they can only open by This Ti a software the full name is TM image and the analysis software uh in order to open the software by some other uh system I would Like to save to a picture file so go
back to here go back to the uh here we need to right click doing when the mouse is here and export data we go to the same location we just saved our picture and we can save this data here and before you click the save you see this T file but please choose CH the second one this one is with the scale bar the first one is without scale bar so please choose the second one and Click save so after that we can see this is a file we just saved and it have the scale
bar there okay so this is how we use camera to save the pictures after save the picture if you want to go to a little higher magnification or you want to go to a little low medication or go to different position please press R1 to go back to the um fare screen and then you can move different location and Even you can go to the different main location for example right now I want to take some uh low Mac image okay for PA like this kind of image when I feel the location is good then
I can start to search leave to the screen so the camera can see the information then check the count again right now the count also a little high So I need to go back adjust the intensity to lower the this bar okay wrong way then open the software again leave the screen and recheck the counts right now you can see the maximum counts here this area it's 600 something like that we just do some quick R average sometimes you'll see it's a little over pretty high but we just do The average here it look like
600 something so that's good then I can acquire to get another picture and in the same we can file save as save the or right now it's already on my folder save the original file and save export data to uh T file select the second one so this is the second pictures you can see this is the first one this is the Second one okay uh I also want to quickly show you show the user how to take high resolution uh images so I go back to the forer screen then I increase medication to uh
close to 630k 500k change intensity to get rid of the noise shrink shrink the Bean okay pretty good let's say The Beam not in the center so I need to Center the beam beam shift and adjust a little height during this H medication to uh re fine adjust the entric height okay then go back to the camera here search lift the Screen and check for the counts right now the counts showing on here I think 12 12 13 1400 is a little bit higher a little high so I need to go back to the here
first first screen and the lower a little intensity so right now the intensity is here I need to lower it a little bit okay so in this magnification I want to use fine okay something like that go back to the Camera okay so right now the maximum number are around here 700 800 that's good less than 1,000 so I can take the picture acquire and the information is on here so uh here it's the room in you can room in you can see the features there this particles it's gold uh it's gold and uh particle
and I think it's uh metal so you can see the metal uh structure here I can also use this one you can say it's Everywhere here here here yeah and this is a good TM image and you can see here it's I can zo in a little more so you can see it here it's 2 nanometer okay so you want to go back right click Auto scale go back to the medication and I can save this picture file save as save the original file and the right click X export data to export to our uh
T file with scale Bar so this is the file I just saved yeah uh this is the Fed window I also can go to the actual size so you can say the metal particles structure okay okay after uh take the pictures after finish take use a camera to take the picture uh before you close this software make sure you saved all this datas we already saved so we can close it and also Another thing to remember before for close the software right now you see here the camera still inserted so please click insert again to
remove the C camera so after that then you can close the software otherwise when the software close this window this box most ini information will be gone you don't know the camera insert it or not fle I Reser insert the camera yellow right now the camera is inserted but if I close this Software oh do I want to save I already saved so you see this box camera information is gone you even don't know it's inserted or not but in fact it's it's already inserted so this is why please remember before you close this software
if the camera is still inserted click again to retract the camera okay then we can close this software for the for the Camera okay well we go back to the for first screen okay and the uh if you want to go to Other Place do some other analysis we can continue working on that one if you finished this sample you would like to get the sample out so uh we opened the V over vacuum overall and also go to the first page you can see right now the the valve is opened and you can see
here so one of the most important is Every time when you insert the holder or remove the holder make sure you closed the valve okay then the next step is you can see here right now this sample XY Z they have some numbers we don't want to remove the sample right now because if you right now they are not Z z0 if you force to remove the holder it may damage the only which may take a few days to two weeks even sometimes three weeks to rebuild the the holder that is take so long so
we would like to you go to the Second page search and reset the holder then we reset everything back to zero Okay so so right now XYZ stage zero and the Volve close so we can start to remove this sample um I will insert uh another holder previous one is a single tiled holder right now I will insert the double tiled holder and demonstrate a sample uh before we insert we make sure the valve is Closed the X and YZ is reset reset holder and also the column vacuum we want to this one smaller than
10 before we insert right now two is very good I inserted the holder in the uh Lo log so it start to pre pump and you can see previously some color change which means different vacuum level and the since this time I'm working with Double T holder so when this says select the holder spacement holder I should select The Double T holder and confirm after confirm since we use Double T holder they have a cable to connect it to the stage so they ask you connect the cable before I can't really physically connect it right
now this is uh not available fit out so I need to connect the cable [Music] first okay when I physically connected the cable you see this one one become to available so I can confirm it so right now everything is good I'm waiting the Pre Pump Air um air lock cycle I still have 2 minutes 20 some seconds then after that I will fully insert the holder in the column so you have like a 30 some second for the uh air loock pump then uh will I will fully insert the holder for double til holder
is inserted right now the column vacuum is eight we wait a Little bit 1 minutes to see we want to make sure this Lumber keep stable or become a little smaller for yeah so right now the column vacuum become to seven that's pretty good so we can start okay the same uh right now I open the valve okay and uh we also loading this file And click your Centric Focus to reset the lens and sunam okay I decrease the medication we want to find our sample our feature okay so this sample uh it's uh uh
the half moon uh sample made by uh du Bean fine Bean system uh so it's like a half circle so you can see this side the right side is vacuum just the uh um left side they have some features There and we can our sample it's on this part so well you mean and and we can go to like uh 2,000 something medication okay okay the sample uh it's a little correct in the middle but uh that's fine we should be still working at that it's the the part should be still okay so I also
adjusted the entric height by Search wer I start wer oh you can see they move out of the screen so that means the uh the height is pretty far away from your Centric I will like to lower a little medication and try this again I want to in the one screen the move in the same time I also can decrease how much degree they're doing right right now it's 8 degree I can move six or even five so I start a rer okay then I use Z to adjust The lumber on here same this is
pressure sensitive if I hold it it will continue so you can see from the Z the continue move for it's a little better so I want to continue in this direction okay I stop right now now because right now our feature is a little to the right so I want to move our features Back and increase a little bit medication basically if they just move in one screen that's good for me okay start the Orit again okay continue okay stop and move to the center okay so you can say the lot top part and the
bottom part they're doing the different move because here have a crack okay but right now we are focus on the Top part so it's good you see this is 500 nomer and right now when I doing the oer they are less than 500 that is uh pretty good so we can stop this one okay so it's already on the entric height okay for this sample I want to demonstrate the STM function uh since TM function I already demonstrated by uh using single T holder I want to use this one to demonstrate the um STM function
so I Find out the the setting file register okay we can use this one STM edx 200 K Spotify I select this file then I click state okay so um we are waiting okay so right now they are they are in the um STM mode so first I would would also want to show you we also can use the Same software tier to operate about the um STM mode so we go to here STM and right now it's already active and uh I can click search uh they can working on that right now the already
selected the high angle agular dark field detector okay so I click search we should see some information of course we need to go to a Little low mag right now the magnation it's a little high okay so this is the sample the feature okay we can use the uh adjust the Z to get to the right Focus adjust a little bit height because STM and TM the the height of it's a little different settings okay I can zoom in a little bit more and move to some features [Music] there Okay so I adjust a little
height for this one you can see you can see the surface clear okay looks like that okay so this is how we can get the pictures and this is the uh hard off image and we can go to a little low you can see the features here uh it's re reversed okay uh and I also want to show you another good software for operate the during the STM mode so I uh stop the Search and I opened the software this software they call it uh vlock so I open it okay so this is the vux
software they also uh designed from uh to operator right now it's in STM mode and uh for the STM our we have four different detector I open this a little bit up so we can take a look details so we have the harduf dark Field 4 dark field 2 and Bright field we have different detector and also from here you can see what is the uh this is their detector layout this is the beam Electron Beam come out come down this is the sample and uh this is our main screen and the camera is under
it I mean the the CCD camera but for the STM mode we use the detector so we have harduf detector dark Field Four you can see the detailed information harduf is high angle angular dark field STM Detector dark Field 4 it's a 4 mm whole radius dark field detector and dark field 2 2 mm hole and also the last one bright field okay so right now the main screen block locked so we only can get a harduf information so let's say I click the harduf then this is hard of information this is the current we
get uh in order to get other information we need to leave to the main screen then we can use other detector for detect the Information so for example right now I open dark Field Four dark field uh dark field two and bright field okay so right now you'll see for this Tre right now no signal because the screen is not open so I press r one to open the scen then we can get the dark field four two bright field and the hard off information however if you pay attention you know the hard off and
the bright field should be reversed this is bright Field image so the heavy materials like Here Pon or some other materials metal should be black other light materials should be um bright so H off should be reversed right so heavy materials should be bright and the light materials should be black but you see dark Field 4 and dark field 2 they more like bright field instead of dark field the reason for that one is because we use the pretty big camera lens in this camera lens we are not get good dark Field image Because the
for that camera lens the bright field too broad so touched the dark field two and four to give the dark field two and four more like bright field instead of dark field we will adjust that one in a little time okay so to adjust this one we just lower the camera lens right now it's 260 so for example I continue lower it 205 160 125 and 98 okay and 77 okay 60 okay so for example on 60 you can see the dark field for the image more like the harda it's not like the bright field
right right you can see the changes so we need to continue right now dark field Q St looks like bread field so we need to continue lower the camera lens Okay okay so right now you see when I use this small camera you can see the dark field two is different with that one so um so they are for the electron be they're not that bright uh broad to touch this one okay uh so um let me uh because right now it's not high magnication I need to zoom in a little bit more to get
more details okay okay for this sample um I right now I uh demonstrate this one this one will be when we choose different uh uh bright Field dark field because some of the informations we see from different detector okay I can rotate this one a little bit okay that's good okay so okay so right now um uh this is hard duck Four duck two and Bradfield for each one of them you also can come to here to adjust the gain and the offset I would Like to because sometimes you say oh this one may be
too black too black I don't see the features some features you can see here some features are not showing so um I would like to use this one this is scope twos and then you can adjust for example right now I want to this bace and this F for this one I can use increase the gain so for example I increase this one I don't want to over the top uh red One and I can l a little bit this one okay in the same time for this one I also can do some adjustment and
here and the bright field okay so when this is um on here right now we use four detectors and we can see some of features some of them uh for example Here they don't showing but we can see pretty here this is trestor and this bright field dark field uh darkfield 2 dark Field 4 and hard off and you can see some of the scenes for example this is the um metal Fielding you can see from the bright field you almost don't see anything information here but in here you can see a little details even
the different structures you can see uh different phas or uh metal uh this is uh different metal materials basically this Is trestor and this is uh v um and here it's uh different other materials so you can see the features different okay um yeah um I will go through this one a little later how we save the picture but uh because we still have need to do some alignment right now just to show here have different detector they can uh when when you have the samples you can open different detector to get the different kind
of feature for the sample Okay so so for the alignment this one right now I stop the search and uh I uh temporarily remove all this kind of detector I put down the for screen okay and I change the medication down uh right now this is being blanked so I active the beam okay the beam is here and uh uh I already find the beam so I don't need to change the medication if the beam it's not here so I need to uh decrease the medication so I can find the beam but right now since
beam is already there so I don't need to um change it medication to find the beam then I go back to high medication like a 630 okay and uh uh right now the beam is so small because the camera lens it's pretty small so I would like to uh increase this one so we can EAS it to align uh do alignment for this mode so I Will go back to the higher l so you can see the beam it's bigger okay okay I would like to uh uh turn off the direct defraction mode refinder the
beam and move the beam to the center by using here they already opened the directory alignment so I use beam shift to move the beam to the center oh right now you can see the beam it's affected by the sample so better I Move the sample a little away okay I go back to here redo it this way so I can easily to move the sample away go to the low ma move the sample away I will go back sample SLE metod okay then I stop this one oh um heart of detector out and uh
uh beam blank un blank the beam okay Turn off the defraction mode go to this uh uh this mode and right now I can see the beam it's here and I can um yeah to adjust so the first adjust right now uh I need to go back to this one the beam go back to the center okay and the uh change the medication to 630k then I use the intensity Leist focus and I adjust the focus to get the hot spot minimize the hot Spot you can say here it's a hot spot I want to
minimize that one okay so close if I continue you can see the hot spot become to Big right so I don't like that I want to continue in other way I want to minimize this one become big something like that okay ijust a little bit more so you can see clearly here okay then we finish this one we done okay then the next step it's rotation Center objective we want To um here when it's click you can see this start to O Okay we use the focus step not the focus focus step to stop this
W okay and after we uh then we want to adjust multi function X and Y to make sure the hot Center is in the center of the beam right now you can say it's on the left side so I adjust this one make sure the uh in the center okay something like that then I click done right now I want to move the Beam back to the to the center so I move it to here yeah sensitive okay um I make it a little bigger okay that's fine uh the next one it's we want to
do the uh make sure this is the circle right now this is okay this side maybe because some part of the sample I still a little close affected so um for the stigma to make it the circle by right Now I think this is okay so I I temporarily ignore this one so I will do the next one it's NP beam um T PPX and pp1 okay this one when you click it you'll see looks like that and you can see we want to the top part here and here come together so I use multif
function to move in this one okay and move our y function to move the tip Together so this X then we do y okay the same thing okay right now you say when I do y they move out of range so I need to lower a little bit medication so I can see the move in here X and Y okay then after that I can use beam shift to move this one to the center okay then I want to repeat the alignment again to make sure this is the Uh settings good so uh I will
do the um intensity Leist focus and I want to make sure the hot spot is the smallest one okay something that's there then rotation Center objective stop the obler by Focus steps okay then I use multif function X and Y you can also say from here multifunction X or Y right now I'm doing those rotation Center X and Y to move the hot spot to the center of the beam You can see it okay looks pretty good I use a high re lotion so in this way can see a little better okay then I redo
Nano beam T PPX and ppy this one looks looks good okay then move the beam shift move back to the center Okay right now the beam looks okay I mean if it's not a circle then you to do The Stig so I just open here Stig but here I should use objective okay to do the stigma so when I click this one you know this is a current settings and also I make you I make a copy right then you can adjust to basically to make the beam wrong fle this is bad right to adjust
as best as you can okay oops okay then I double check is this One better or this one better right yeah I think both seeds pretty good yeah okay so this is stigma okay when we finish this one alignment uh then the the uh we need to do this one to get the um the sttm image yeah so I go back to here and I go to the electro mode okay then I go back to the vux software okay I go back reopen the hard Off detector and go back to the deflection mode okay so
this is a sample right and uh we can zoom in I think for this this is a semiconductor sample on the top part should be uh uh amorphus platin deposition uh should be ion beam or or Electron Beam induced platin depilation this is a morace so we can uh take a look on this area here for For okay so this is the U plattin there area um we also can stop the search and move this one in the platum area then I'm putting the p okay and we can see this is um I use high
contrast this is the we're adjusting the RO G so uh for here I use high resolution can see more Clearly so uh I adjust first adjust the focus so you can see like a zo out right and come in the infinity then room out you can see I can see the feature surface feature but I want to go to Infinity like here roughly like this this location otherwise if I continue it will be room out okay you can see it here okay so I move to something like here like here then I will start to
do the uh s so I open this one and I use The condenser okay so select the condenser and this is the current settings I make copy copy one to two before I adjust anything so right now I adjust adjust one of them so for example right now ijust the condenser stigma X I change the X you can see when I change it something bad you can say oh I see a line or something like here that is Stig right so I will do a another way if I do too much you will start to
see the Stig Like here to show again so I want to just in the middle to don't see the nine features there okay something like here then I will do another multifunction y do the same thing I see like this one I say the line horizontal then I do other way I say like the line vertical right so I just want to be in the center the area okay I redo make sure X is good then I uh redo the focus so stickat and focus Back and forth several times to get this one so for
the Focus right now the steps it's a little high right now it's previously it's four I right now I make it a three and uh I want to adjust it a little for the focus okay this one let zo out zoom in room out the maximum it's here and uh then from here I adjust the stigment Here right here okay and move a little bit off find a l g location for for for okay something like that okay like a eyeball or something that is adjustment for here so when they finish these Steps I can
go back to here and I can uh start to searching again yeah so you can see previously when I adjusted this location and this location the beam makes some damages okay so let me see about this location how the is okay this location it's okay looks okay you don't see the line and it's like yeah like the center like The eyeball here and outside looks good okay so right now I can start to search again in we adjust the stigma so I can go to the features I would like to show you right now this
is a hard off detector so like I say if it's bright that means the heavy materials if it's um darker that is the um uh bright materials so let's go to the transistor here here e And we can open here adjust brightness contrast make sure it's good okay then so here it's search and you can see search and uh this one and this one is two settings for AC chire for example this one is good then I want to take some pictures I just click here they will finish the scin and get the picture for
the hot do picture on this location okay and in the same and for this one You also can save save the file processing it's this one so you can say this just the image we saved so we can export this one to uh export the image and we can change to the folder the same folder we save the other pictures and we can choose this folder so this is the pictures it's this ISS the date uh STM what magnification ad picture so we can export to this one okay so after that one we can open
here This is our hu picture yeah um for STM one of the features you can use different detector for different functions for example this is a uh harda picture and also we can get the different one for example if we want to go to the bright field we also can get the same bright field but for the bright field detector we would like to uh open the main screen so in this way the image can go to the bright field and we can Get [Music] the pictures here okay so I can come to here also
save the picture so they will take two pictures um hot off and the bright field okay and you can export uh this one I think we already did that one seminar and we also can export here on that folder so this bright Field image after Uh export it we can take a look so you'll say this is the dark field bright field okay image um so when using STM they can function they can get a different features they can have different detectors another thing very useful it's for uh they can also do the edx so
the edx function is here and uh um example let's take a look we get one edx Okay give me one second right now I think the beam it's uh blank so low beam so I need to open the beam then I should be able to take some edx signal okay so you can see they get some edx signal there and uh um we can move this one to different location we can say it should be say different signal things there if I move to Here okay different signal different location okay you can say different materials
coming up out okay it's uh we can do some matching so for example like uh I set up something here I'm working on this window okay and also we can set some place here so this drift area the drift control and uh uh for here we can say say like uh derail time right now I set the derail time uh 16 Second uh take little long I set this one maybe a little small so we can save some time okay so this one take 3 minutes that's good 4 minutes 7 minutes okay this one should
be work okay so I set up doing a mapping on this area and this one is a drift area so after that I can start during the edx they scan and they uh get the reference area and also they are um uh doing the edx Okay so I open the accur so right now they doing this one so we can choose different materials for example silicon we doing the mapping and the oxygen this is previously we find out and also uh I think this part it's nitrogen okay and the top part I think uh uh
you that's Cobo okay so right now I'm waiting uh they They're doing uh continous scan and this is the processing window so they can see the update information here and uh uh also we are doing the equalization we can see from here there scan and this is the cross correction basically reference and that one if some changes they will be corrected okay uh I can see right now the sample it's a little drift to the right looks like yeah maybe I need to change some Parameter but anyway we can get some idea from there [Music]
okay I think I can stop here I think they have some drift the corrections cannot uh fix it so it's a little drift to the right but we still can get the useful information from here uh maybe the sometimes the drift just because the sample just put in the chamber Uh sometimes need a little time to make it stable okay but uh anyway from here you can see the information so this is hard of information and I lay out with oxygen so you can see this part it's oxygen and outside and silicon almost everywhere and
this is a cobble I think they have some drift there metal and this is nitrogen center part center part so this is how we do the mapping I will uh try to redo another one because I think this One is a little drift so I will say yeah you can see it's drift to the right I will um I will wait a little time maybe do this again to avoid the drift right now I think it's better than previously yeah and uh I also can lower this a little more okay smaller the real time maybe
they Don't Drift that much okay so looks good I will try Again for for still a little bit drift can sa from here maybe need to find a larger location this location maybe not that stable okay cobal silicon oxygen nitrogen yeah so should um I think most Likely maybe the be either the sample not that stable right now need to wait a little bit time because when insert the new sample some times the sample and the inside the max scope maybe the temperature different so they have some seral drift or um I use a little
higher beam current make the sample not that stable need to use lower beam current anyway you also can see this sample have a little crack there maybe also make that one not that Stable okay so this is about the edx and uh um take some pictures for for the sample so let's go to uh a little uh NOC let's take a look other uh location of the sample for for so bright field and the dark field right now I adjust some gain and offset I don't want to Satur it this one Looks good adjust to
the breath field for okay no Mac um hot off image and bright field image and we can take the photo here check the parameter that's good take a photo here okay and also can save the picture Export export here okay so this one should be bright field previous one is hard off okay okay so one finish here weish finish this one so uh uh we would like to uh remove the detector especially the um especially for the uh bright field because the bright field detector you can see from here they will block the beam so
if you this one you don't remove uh retract the camera it will be broken For the TM mod to the C is the camera so get this one and the hard off you also can remove that one so then after that uh we go back to go back to the TM mode which set up and we find the features TM okay spot five then set it will be uh go to the TM mode then we can take a quick look make sure everything is good we put it back the be uh the screen and uh
be is blank I'm BR the Be okay so this is the picture Okay so that's good so right now we finished the uh this holder for this sample thing I want to uh demonstrate it's about the deflection so this is the uh I use this this sample as a example so uh I go to a little low Ma you can see here this is the uh silicon uh this side is a divice and here it's a silicon so I just go to one of the red location and I click the Defraction and I can see
some patterns there I can adjust change the by Mouse change it to to say the kakuchi N9 you can see there and right now you can see the major kakuchi n you can say here to be on this location so uh when we want to get the high magnication uh TM image or sometimes STM image we want to the be sue the room axis and right now from this kuchin line you can see here it's a room access so I want to right now it's a little off I Want to move move this one to
the center so you can see this one for example right now I make it worse it's on here you can see I want to move this one to the center so I use since this is the um double til holder so I can use Alpha and beta to move if a single TI holder we can only REM move one of them so right now I use Alpha I move this one to here then I use B beta to move from here from here to the top to the center Okay um SLE a little I move
okay good I think before they have a jump maybe it's something broken there for sample you know it's some crack there so right now this part is good I move this one to the center and uh I I I use deflection alignment I can move this one to the center for for For for for okay so this one right now it's very close to the room axis you can see this this is the center beam and we can have the insert the beam stopper to be the center and then we can take the pictures for
this one yeah so uh This Is How We Do the defraction to get the room access uh for the sample okay so we go back to there And uh when we finish this sample we REM use this one to remove the beam blanket okay then we can come to here and stop the um close the wall and reset the holder then we can remove the sample reset the holder you say RHA X Y uh x y z need all be set to Z